• Group-by: Investigate patterns in the data by grouping rows based on user-defined
conditions. Use of this query is described in more detail the Core Standard Platform
Software Operations Manual (SOM).
• Aggregate: Investigate patterns in the data by aggregating rows based on user-
defined conditions. Use of this query is described in more detail the Core Standard
Platform Software Operations Manual (SOM).
11.7 Diagnostics Screen
Figure 15. Sample Diagnostics Screen
The Diagnostics screen displays real-time results of system diagnostics including:
• CPU: Detailed information about PLC CPU diagnostics.
• CPU Major Fault: Details about the last or current CPU Major Fault. If the Major Fault Active
text is not bright red, then there is no major fault active.
• Heartbeats: Information about the bidirectional heartbeats between the PLC and HMI. The
length of the trend showing the heartbeat signal can be adjusted from this screen to view a
longer-term period.
• HMI Diagnostics: Displays performance and status information about the software
components for the HMI. All items should either be On Scan or Connected.
Clicking on the names of the components will display a detailed faceplate with diagnostics
and engineering functions. All users can view the faceplate, only users with engineering-level
security can execute functions.
• Indicators: The current status of miscellaneous indicators for the system that don’t fit on the
process screen. Note that the instruments and layout of this section of the Diagnostics
Screen may not represent your installed system as some functions may not be available.
o Alarm Lights: The current state of the alarm lights (stack lights) as well as the ability
to test the outputs for each light. This is only available on systems with system
alerts. Alarm Lights testing can only be performed by users with Supervisor level
security.
o Alarm Horn: Displays the current state of the alarm horn as well as the ability to test
the outputs for each horn. This is only available on systems with system alerts.
Alarm Horn testing can only be performed by users with Supervisor level security.