Page #26 background image
Rinstrum - 5200 Communications Manual Rev 1.2
Page 26 0025-605-120
2.4.15. FOP: Force Output
Use this command to force an output either On or Off. This is only available for
outputs that are setup with no other function. Use the query to obtain the state of
the four outputs.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
4 - Immediately No
Parameter Details
Parameter Description Range Default Required
1 Output 1 0
1
Output 1 OFF
Output 1 ON
0 No
2 Output 2 0
1
Output 2 OFF
Output 2 ON
0 No
3 Output 3 0
1
Output 3 OFF
Output 3 ON
0 No
4 Output 4 0
1
Output 4 OFF
Output 4 ON
0 No
Example
Command Response Details
S01; Select Instrument 1
FOP?; 0,0,0,1 CRLF Output 4 is ON. All others are OFF.
FOP,,1; 0 CRLF Drive output 3 ON.
FOP,,0; 0 CRLF Drive output 3 OFF.