Rinstrum - 5200 Communications Manual Rev 1.2
Page 32 0025-605-120
2.4.21. LIC: Linearisation
This command gives access to the multi-point linearisation functions. A special
query, LIC?;, is available to verify the linearisation correction.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
2 With TDD1 Immediately Yes
Command Parameter Details
Parameter Description Range Default Required
1 Linearisation Point 1 to 5 1 No
2 Test Weight 0.999999 - No
Query Parameter Details
Parameter Description Range
1 Percentage of fullscale weight reading –100 to 100
2 Correction (in weight unit x10) –100000 to 100000
To clear one of the linearisation points leave the test weight value off.
Example
Command Response Details
S01; Select Instrument 1
LIC1; 0 CRLF Clear linearisation point 1
LIC?1; 0,0 CRLF No correction for point 1
MSV?; 120.5 CRLF Query measured weight value
LIC1,1200; 0 CRLF Correct for the current test weight of 120.0kg
(Note that weight is sent without any decimal
point. So 400.0 kg is sent as 4000 not 400.0)
LIC?1; 24,–50 CRLF Current linearisation is approximately –5.0 kg at
24% of fullscale reading.
TDD1; 0 CRLF Save Setting