Rinstrum - 5200 Communications Manual Rev 1.2
Page 30 0025-605-120
2.4.19. LBT: Button lock settings
Sets the operation status of each of the four front panel buttons.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
2 With TDD1 Immediately No
Parameter Details
Parameter Description Range Default Required
1 Button 0
1
2
3
Zero
Tare
Gross/net
Print
0 No
2 Operation 0
1
2
Lock
Normal
No add to totals
(print key only)
1 No
Note: Operation 2 only applies to the <PRINT> key. In this mode, printing is
possible but the current weight will not be added to the product totals. This is a
useful feature for checkweighing where totalising is done automatically.
Example
Command Response Details
S01; Select Instrument 1
LBT?0 1 CRLF Zero button is set to normal operation
LBT0,0; 0 CRLF Block operation of the Zero button
TDD1; 0 CRLF Save Setting