EasyManuals Logo
Home>Rinstrum>Measuring Instruments>5200

Rinstrum 5200 User Manual

Rinstrum 5200
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
Rinstrum - 5200 Communications Manual Rev 1.2
0025-605-120 Page 43
2.4.31. PST: Print strings
Set the print strings. Print strings include the header, footer and the three user
strings. Refer to the 5200 Digital Indicator Reference Manual for more
information.
General Details
No. of Parameters Save Changes Changes Used Increment Trade Counter
2 With TDD1 immediately No
Parameter Details
Parameter Description Range Default Required
1 String Token 172 to 177 - Yes
2 String String. Length <= 230
characters
(String IDs172 to 174:
Length <= 6 characters)
“” No
Example
Command Response Details
S01; Select Instrument 1
PST?175; “” CRLF Query the print header
PST?175,"McFarlan Meats\013
\010Ph: 3375 3120";
0 CRLF Set the print header to:
McFarlan Meats
Ph: 3375 3120
TDD1; 0 CRLF Save the changes