Chapter 8 Control, Inspection, Exchange and Cleaning
HRR Series 8.1 Quality Control of Circulating Fluid and Facility Water
Chapter 8 Control, Inspection,
Exchange and Cleaning
8.1 Quality Control of Circulating Fluid and Facility
Water
Table 8-1 Quality criteria for clean water (tap water)
Electric conductance
(at 25
o
C)
Acid consumption
(at pH 4.8)
Quoted from the standard “JRA-GL-02-1994”, The Japan Refrigeration and Air Conditioning Industry Association.
Use specified fluids only. If other fluids are used, they may damage the
product, causing fluid leakage, or result in hazards such as electric
shock or leakage of electricity.
When using clear water (tap water), ensure that it satisfies the water
quality criteria shown in the table below.
Replace the circulating fluid and/or the facility water if any problems
are found in the regular check. Even if no problems are found, some of
the water in the tank evaporates and impurity concentration in the
circulating fluid increases. Replace the circulating fluid on the tank
once in every 3 months. Refer to the section "8.2 Inspection,Exchange,
Cleaning" for regular inspection.