1Ć6
   !   
 
Main Seconds/Division Range 20 ns/div to 0.5 s/div.
The seconds/division ranges from 20 ns/div to 0.5 s/div in a 1-2-5
sequence of settings. The X10 magnifier extends the maximum
sweep speed to 2 ns/div.
Delayed Seconds/Division Range 20 ns/div to 5 ms/div.
The seconds/division ranges from 20 ns/div to 5 ms/div in a 1-2-5
sequence of settings. The X10 magnifier extends the maximum
sweep speed to 2 ns/div.
Sweep Length >10 divisions.
Delay Control Range 0.15 division to 10 times the Main seconds/division setting. The
maximum value can not exceed the end of the Main sweep.
Delta Time Control Range 0 to 10 divisions to the right of setting of the delay control, but maxiĆ
mum value does not exceed end of the Main sweep.
   !  
 
Trigger Level or Threshold Ranges The ranges are as follows:
 
Internal ±15 divisions from center of screen
with ground at center graticule.
External 1X ±1.5 V.
External 10X ±15 V.
Line ±400 V.
   !   
 
Slope Selection Slope selection must match the polarity of the sync (i.e., for negaĆ
tive going sync, negative slope must be selected).
A default slope selection for TV trigger modes can be entered in
the scope configuration menu.