4Ć7
4. Verify that the main time base is operational, confirming the following
statements are true.
H One period of the square wave is about five horizontal divisions.
H Rotating the  control clockwise expands the waveform
(more horizontal divisions per period of waveform); rotating it counĆ
terclockwise contracts the waveform.
H Setting the sec/div scale to 1 ms displays approximately one period
of waveform per horizontal division.
H Rotating the horizontal  control moves the waveform left
and right onĆscreen.
H Pressing the  button changes the sec/div scale from 1 ms to
100 ms and one period of the square wave is about ten horizontal
divisions. Return to a nonĆmagnified display by pressing the 
button again.
5. Press the   button and set  to &.
6. Set the delayed sec/div scale to 500 ms.
7. Press the   button and set  to +&) *(.
8. Press the   button.
9. Select  from the main menu and vertically position the delayed
time base below the main time base using the General Purpose Knob.
10. Select  from the main menu.
11. Verify that the delayed time base is operational, confirming the following
statements are true:
H The main sweep has an intensified zone of approximately 5.5 horiĆ
zontal divisions. Adjust the  level if necessary to view the
entire intensified zone.
H Rotating the General Purpose Knob clockwise moves the intensified
zone to the right onĆscreen. Position the intensified portion to the
center of the screen.
12. Adjust the delayed sec/div scale to 200 ms.
13. Set  to  and confirm that only the delayed sec/div scale is
displayed and the period of the squareĆwave signal is about five horizonĆ
tal divisions.
14. Set  to & and  to .
("- *! (" ( -)*%
1. Press the   button until the readout indicates channel 1
is the only selected channel.
2. Press the  button.