Reference
GPIB Guid elines. Follow these guidelin es when you connect y our
oscilloscope to a GP I B network:
Turn off the oscilloscope and all external devices before connecting
the oscilloscope to the GPIB netw ork.
Assign a unique device address to the oscilloscope. Two d evices
cannot share the same device address.
Turn on at least two-thirds of the GPIB devices while using the
network.
Calibration System
Use the System Cal menu to access these functions.
Bottom Side Description
System Cal
Signal Path Compensates the signal paths to
obtain best measurement accuracy.
Factory Cal
Used to calibrate the oscilloscope.
This is a service function only.
Notify After Hours
of Operation
Sets the number of hours of
operation before notifying you that
a calibration is due.
Cal Due
Control
Notify After Years
of Elapsed Time
Sets the number of years before
notifying you that a calibration is
due.
Key Points
Signal Path Compensation. For maximum accuracy at any time, run the
SPC routine just before taking critical measurements. To meet accuracy
specifications, run the routine if the ambient temperature changes by
10 °C (18 °F) or more.
Before running the routine, disconnect any probes or cables from the
channel inputs. Then, push the Signal Path and th e OK Compensate
Signal Path screen buttons to confirm that you are ready to proceed. The
routine takes several minutes to com plete. (See page 4.)
Factory Cal. Service personnel use these functions to calibrate t he
oscilloscope internal voltage references using external sources. Refer to
138 TDS3000C Series Oscilloscope User Manual