Remote Interfaces
TP04100A Interface
&
Applications
Rev
A:
0611
5/98
,,&?,*
CORPORATION
v
TEMPTRoN'C
4-8.2
Device Specific Commands and Queries
ADMD Set the air-to-DUT maximum difference
ADMD nnn
--
where
nnn
is 10 to 225 degrees C in 1 degree increments. Default value: 225
ADMD? Read the air-to-DUT maximum difference
AUXC?
Read the auxiliary conditions register
bit 0
--
na
2
--
dut mode
3
--
flow on
4
--
na
5
--
na
6
--
manual mode=O, program mode=l
7-na
CFIL
Copy setup(0) to setup(n)
-
where n is the #(I-9)
CLER
Clear the device-specific (reported by EROR?) errors
CYCC
Set the ramplsoak cycle count for the current mode. Default value: 25
CYCC nnnn
--
where nnnn is the
#
(1
-
9999) of cycles to do
CYCC?
Read the number of ramplsoak cycles to do
CYCL Startlstop ramplsoak cycling
CYCL 1
--
starts
CYCL 0
--
stops
NOTE:
When all cycles have been completed or when cycling was stopped on failure, it is necessary to
send a CYCL 0 command to reset the system.
CYCL?
Read the Ramplsoak cycle number
(current value
if
cycling, last value if not)
Page 4-12