TEMPTRONIC
CSRPOR4TlOh
DSNS
DSNS?
DTNN
DTNN?
DUTC
DUTC?
DUTM
DUTM?
EROR?
FLOW
Rev A: 0611 5198
Set the DUT sensor type. Default value: 0
DSNS n
--
where n is 0-2
0
--
no DUT sensor connected
1
--
type
K
thermocouple
2
--
type T thermocouple
Read the DUT sensor type
Turn DUT Tuning on or off
DTNN
0
--
off
DTNN 1
--
on
Read the tuning status
Remote Interfaces
TP04100A Interface
&
Applications
Set the device thermal constant. Default value: 100
DUTC nnn
--
where nnn is nominally 100 but can range from 20-500.
A
higher number
corresponds to a higher mass device.
Read the device thermal constant
Turn DUT mode on or off
DUTM 0
--
off (air control)
DUTM 1
--
on (actual DUT control)
Read the DUT mode
Read the device-specific error register (16 bits)
bit 0
--
overheat
1
--
air open loop
2
--
air sensor open
3
--
low flow
4
--
na
5
--
na
6
--
na
7
--
system error
8
--
na
9
--
No line sense (line power sense error)
10
--
na
11
--
nvram error (non volatile ram error)
12
--
bram failure (battery
ram
failure)
13
--
na
14
--
dut sensor open
15
--
na
Turn
the flow on or off
FLOW 1
--
on
FLOW
0
--
off
Page
4-
13
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Temptronic ThermoStream TP04100A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Temptronic ThermoStream TP04100A and is the answer not in the manual?

Temptronic ThermoStream TP04100A Specifications

General IconGeneral
BrandTemptronic
ModelThermoStream TP04100A
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals