5-40 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
2. Disconnect the Sheath Flow + tubing from the DMA and connect it to
a reference flowmeter with a flow accuracy of ± 2% or better over the
range of 2 to 30 L/min, such as a Gilibrator
®
meter.
3. Connect the Sheath Flow port on the classifier to the Sheath Flow
port on the DMA if not already connected.
4. From the classifier Setup screen navigate to Calibration > User
Calibration. Select the Sheath Flow tab.
Figure 5-24
Electrostatic Classifier User Calibration ScreenSheath Flow Tab