6-32 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
Retur n i n g the Ele c t r ostatic C l a ssifier for
Servic e
The Electrostatic Classifier should be serviced at the factory facility
annually for maintenance of the blower, filters, flowmeter, and other
sensors, and DMAs. An ISO 15900-compliant factory calibration will be
performed at this time. Before returning the classifier to TSI for service,
visit our website at http://rma.tsi.com or call TSI at 1-800-874-2811 (USA)
or 001 (651) 490-2811 for specific return instructions. When you call, have
the following information ready for the Customer Service representative:
Instrument model number
Instrument serial number
A purchase order number (unless under warranty)
A billing address
A shipping address
TSI recommends that you keep the original packaging of the classifier for
use whenever the instrument is shipped, including when it is returned to
TSI for service. Always place protective caps over the inlets/outlets to
prevent debris from entering the instrument. If you no longer have the
original packing material, place the instrument inside a plastic bag to
protect it, and then package with at least 5" (13 cm) of shock
absorbing/packaging material around all sides of the instrument.
C a u t i o n
Do not ship the Model 3082 Electrostatic Classifier with the lead
shielding, the Neutralizer, or the DMA installed.
Lead shieldingDo not return the classifier with lead shielding
installed, as this may cause damage to the classifier during shipment.
3077/3077A NeutralizerFollow safety regulations for correct
shipment of radioactive sources. Consult the appropriate Neutralizer
manual for details.
3088 NeutralizerThe 3088 Neutralizer can be serviced by TSI, but
should not be installed during shipment to avoid damage.
DMAThe 3081A, 3085A, and 3086 DMAs can be serviced by TSI,
but should not be installed during shipment to avoid damage.
See Chapter 3, "Moving and Shipping the Electrostatic Classifier" for
detailed instructions.

Table of Contents

Other manuals for TSI Instruments SMPS 3938

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments SMPS 3938 and is the answer not in the manual?

TSI Instruments SMPS 3938 Specifications

General IconGeneral
BrandTSI Instruments
ModelSMPS 3938
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals