EasyManuals Logo
Home>TSI Instruments>Laboratory Equipment>SMPS 3938

TSI Instruments SMPS 3938 User Manual

TSI Instruments SMPS 3938
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #144 background imageLoading...
Page #144 background image
6-2 Electrostatic Classifier Model 3082 and SMPS Spectrometer Model 3938
Table 6-2
Electrostatic Classifier Maintenance Schedule
Hours of
Operation
Maintenance Task
5 to 50
Clean the impactor. (See Cleaning the Impactor, below)
4000
Clean the mobility analyzer center rod and outer electrode in the
Electrostatic Classifier (see Cleaning the DMA Electrodes, below)
4000
Perform O-ring maintenance (see Replacing Impactor O-rings and
Greasing O-rings, below)
8000
Replace the filter cartridges in the Electrostatic Classifier (See
Replacing the Inline Filters, below)
8000
Clean the DMA Dacron screen (see Cleaning/Changing the DMA
Dacron Screen, below)
8000
Clean the neutralizer (See separate neutralizer manuals)
8000
Calibrate the classifier sheath flow (see Chapter 5, Performing a
User Calibration).
8000
Replace CPC filter cartridges (See separate CPC manuals)
Recommended Cleaning Solutions
The external surfaces of the 3082 may require cleaning time-to-time.
Isopropanol or soapy water can be used to clean the classifier (display,
chassis, impactor, DMA). Dampen a cloth with cleaning solution to apply.
In order to prevent dripping or overspray into instrument, do not spray on.
Do not use any abrasive materials, aromatic solvents (toluene, xylene), or
ketonic solvents (ketone or acetone) as these may cause damage to the
display and/or the painted chassis.
Cleaning the Impactor
The primary purpose of the impactor is removal of particles above a
specific size. Over time, particle loading on the impaction plate influences
the pressure drop across the nozzle and; therefore, particle measurements
due to particle re-entrainment. The impactor must; therefore, be cleaned at
intervals of 5 to 50 hrs, depending upon the inlet aerosol particle
concentration.
To clean the impactor, follow these instructions:
1. Grasp the impactor, twist counterclockwise, and pull it out of the
aerosol inlet.

Table of Contents

Other manuals for TSI Instruments SMPS 3938

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments SMPS 3938 and is the answer not in the manual?

TSI Instruments SMPS 3938 Specifications

General IconGeneral
BrandTSI Instruments
ModelSMPS 3938
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals