Unpacking and Setting Up 2-13
Installing the Model 3088 Neutralizer
W A R N I N G
The Model 3088 Advanced Aerosol Neutralizer contains an x-ray
source, which may be subject to local regulations. During normal
operation, you will not be exposed to x-ray radiation. Do not detach the
x-ray source from the neutralizer tube.
C a u t i o n
When installed, the neutralizer completes a flow path inside the
classifier. If you operate the classifier with an external neutralizer or
without any neutralizer, the provided impactors cannot be used, and
aerosol flow rate cannot be measured by the classifier.
To install the Model 3088 Advanced Aerosol Neutralizer, follow these
instructions:
1. If the flag assembly has already been removed, continue with step 3. If
the flag assembly has not been removed, using a Phillips-head
screwdriver, remove the two screws securing the flag assembly and
save for reuse.
2. Lift out the flag assembly by placing your fingers in the depressions on
either side and pulling up on the Polydisperse Flow outlet port. Save
the flag assembly for reuse.
3. Remove the black vinyl end caps from the Model 3088 Neutralizer.

Table of Contents

Other manuals for TSI Instruments SMPS 3938

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments SMPS 3938 and is the answer not in the manual?

TSI Instruments SMPS 3938 Specifications

General IconGeneral
BrandTSI Instruments
ModelSMPS 3938
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals