EasyManuals Logo
Home>TSI Instruments>Laboratory Equipment>SMPS 3938

TSI Instruments SMPS 3938 User Manual

TSI Instruments SMPS 3938
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #147 background imageLoading...
Page #147 background image
Maintenance, Service, and Troubleshooting 6-5
5. Remove the thicker, 1-207 face-seal O-ring (6006680) from the bottom
face of the impaction plate.
6. Remove replacement O-rings and vacuum grease from the 3082
Accessory Kit. Follow the instructions in the Greasing O-rings section
below and smear a small amount of vacuum grease onto the
replacement O-rings.
C a u t i o n
The 1-207 O-ring (6006680) is made of conductive material to ensure
that the impactor is grounded to the classifier chassis. It should not be
replaced with a substitute 1-207 O-ring.
If you set the O-rings on a dirty surface, they will pick up debris. Perform
a visual inspection and remove debris with a clean wipe.
7. Slot new O-rings into the correct grooves in the impactor body.
To replace the 1-025 O-ring (6006389) from the classifier inlet, follow
these instructions:
1. If an impactor or inlet adapter is installed in the inlet of the classifier,
remove it.
2. Locate the O-ring.
Note: It may be helpful to use a light.

Table of Contents

Other manuals for TSI Instruments SMPS 3938

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments SMPS 3938 and is the answer not in the manual?

TSI Instruments SMPS 3938 Specifications

General IconGeneral
BrandTSI Instruments
ModelSMPS 3938
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals