6. Display speed is slow after average sampling
time is enabled :
If average sampling is carried out for more
than 32 times, the display speed will drop. This is
normal.
You can reduce the intervals of average
sampling.
7. Trapezoidal waveform display :
This is normal. The reason is possibly
horizontal time base range is too low. You can
improve horizontal resolution and enhance the
display quality by increasing the horizontal time
base.
If the display type is vector, the connection
between sampling dots may cause a trapezoidal
waveform. Set the display type to dot to solve
this problem.
Chapter 5 Using the USB update
program
The USB upgrade program makes upgrade easier and
more flexible. To use this function follow the steps
below :
1. Check the model name, software and hardware
versions of your oscilloscope in Table 2-26.
2. Download the highest version of the update
program document that matches your model,
software and hardware versions online, and store
it on the USB device (preferably a USB device
recommended by us).
3. Power off the oscilloscope then insert the USB
device. Then power on the oscilloscope again.
The prompt Please press [ ] to continue or
[ ] to cancel appears on the screen as shown
in the figure on
the right
,
.
F5
F1
98
UTD4000 Four-channel
User Manual