EasyManuals Logo
Home>Xo>Dental equipment>FLEX

Xo FLEX Installation Instructions

Xo FLEX
59 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
XO FLEX Installation instructions Installation requirements
8
Incoming water must be in accordance with local drinking water regulations.
If the unit is not supplied with “backflow water prevention”, an external backflow
prevention with an air gap of at least 20 mm is required.
If the incoming water pressure exceeds 6 Bar, a reduction valve must be mounted in
front of the unit.
To ensure correct operation of the unit, the compressor must be an oil-free type and
fitted with an air dryer.
If the incoming air pressure exceeds 7 Bar, an appropriate reduction valve must be
fitted.
2.2 DIMENSIONS OF OPERATORY
XO FLEX is intended to be permanently installed in a dental operatory at least 220 cm wide and
360 cm long
1
see Figure 1. Please also see the installation drawing in a scale of 1:20 (YB-231)
and the installation plan in a scale of 1:1 (YB-235).
Please check drawings YB-231 and YB-235 carefully before finalizing the placement of
the XO FLEX unit in the operatory.
CEILING HEIGHT
The ceiling height of the room should preferably be 255 cm for easy installation of
operating light arm minimum height is 245 cm!
OPERATING AND TRANSPORT CONDITIONS
Table 2 – Operating and transport conditions
Condition
Operation
Transport and storage
Temperature: +15
o
C – +35
o
C -40
o
C – +70
o
C
1
Note: This dimension depends on the depth of the cabinets in the example shown in Figure 1 the cabinets are 50 cm deep

Table of Contents

Other manuals for Xo FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xo FLEX and is the answer not in the manual?

Xo FLEX Specifications

General IconGeneral
BrandXo
ModelFLEX
CategoryDental equipment
LanguageEnglish

Related product manuals