-42-
(7) FC connector
FC connector of the fiber is connected here. Please firmly connect the fiber by
adjusting to the guide key and tightly screw. Please use supplied FC connector cover
for a dust cover in case of storing or fiber exchange.
(8) Fiber guard
Guard to protect fiber from mechanical damage.
(9) C-mount
Camera port. Please use supplied C-mount adapter (D8953CF) to set a camera on
this port, and fix with set screws.
(10) C-mount adapter(D8953CF)
(11) Port switching knob (manual)
Knob to switch light path between 1
st
camera port (pushed) and 2
nd
camera port
(pulled). In the case of A2 and A3 models, no manual knob is provided since they
have motorized switching system.
(12) Dichroic mirror block cover
This cover will be removed when to set or exchange the dichroic mirror block inside.
If you open this cover, Class 3B laser beam could be emitted. Warning sign to class
3B laser radiation to advise to avoid exposure to the beam is attached. Please ask for
the advice from your support service when you wish to exchange the DM block.
(13) Light path selecting block cover
Light path selecting block to set either a dichroic mirror or reflection mirror inside.
If you open this cover, Class 3B laser beam could be emitted. Warning sign for class
3B laser radiation to ask avoiding exposure to the beam is attached.. Please ask for
the advice from your support service when you wish to exchange the mirrors.
(14) BA (Barrier) filter block
Barrier filter block will be set here. When you remove this block for filter exchange,
please turn off the power for laser safety.
(14)-1 Fixing screw for setting the BA filter block or filter wheel unit
(15) Camera port support
To adjust height and support weight when CSU-X1 is installed at the side port of
microscope.