AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 133
Limas21 HW: Gas analyzer start-up
Gas analyzer start-up
1 Turn on the gas analyzer power supply.
2 Wait for the warm-up phase to end. Wait at least 2 hours until the tem-
perature of the heated sample gas line has stabilized at 180 °C.
3 Purge the complete sample gas feed path (sample gas line and gas ana-
lyzer) for at least 1 h with clean and dust-free ambient air.
4 Turn on the sample gas supply.