AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 245
Fidas24: Response factor and other relevant variables
Response factor
Definition
ionconcentrat
display value Measured
factor Response =
or
factor Response
display value Measured
ionConcentrat =
The response factor of propane (C
3
H
8
) is equal to 1.00 as defined.
Response factors for Fidas24 analyzer module
Sample component Response factor
1)
Toluol C
7
H
8
0.95
Chlorbenzene C
6
H
5
Cl 0.95
p-Xylol C
8
H
10
0.92
Benzol C
6
H
6
0.99
Ethylbenzene C
8
H
10
0.92
Propane C
3
H
8
1.00
n-Hexane C
6
H
14
0.97
n-Octane C
8
H
18
0.93
iso-Octane C
8
H
18
1.04
Trichlorethylene C
2
HCl
3
0.96
Tetrachlorethylene C
2
Cl
4
1.00
Ethane C
2
H
6
1.01
Butane C
4
H
10
0.97
Methanol CH
3
OH 0.74
Butanol C
4
H
9
OH 0.83
Acetic acid C
2
H
4
O
2
0.52
Dichlormethane CH
2
Cl
2
1.00
Methane CH
4
1.14
1) Measurement of the components in synthetic air
NOTE
The response factors for an individual analyzer module may differ slightly
from the values indicated in the table.