AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | OI/AWT420-EN REV. B
51
2-electrode conductivity – dual input calculated values setup
Menu Comment Default
Calculation Type Calculations are performed using the inputs from both sensors.
Select the required calculation from the following options:


No Calculation
Inferred pH (NaOH)

conductivity readings.
Note: The temperature compensation characteristic TC Curve for signal B should be set to NaOH.
Inferred pH (NaOH+NaCl)

conductivity readings.
Note: The temperature compensation characteristic TC Curve for signal A should be set to NaCl.
Note: The temperature compensation characteristic TC Curve for signal B should be set to NaOH.
Inferred pH (NH3)

conductivity readings.
Note: The temperature compensation characteristic TC Curve for signal B should be set to .
Inferred pH (NH3+NaCl)

conductivity readings.
Note: The temperature compensation characteristic TC Curve for signal A should be set to NaCl.
Note: The temperature compensation characteristic TC Curve for signal B should be set to .
Difference
Calculates the difference between the two conductivity inputs:

Ratio
Calculates the ratio of the two conductivity inputs:

B
A
% Passage
Calculates the amount of conductivity as a percentage that passes through the cation exchange unit:

A
B
x 1
00
% Rejection
Calculates the amount of conductivity as a percentage that is absorbed in the cation exchange unit:

A
B
x 1
00
1
(
(
Note.
Before Cation Limit Set the required before-cation conductivity limit, between:
 
 
 
 
N/A
After Cation Limit Set the required before-cation conductivity limit, between:
 
 
 
 
N/A
pH Range View the measuring range for the selected Inferred pH calculation
 
 
 
 
N/A
Signal Arrangement Set the signal arrangement:

Note: For inferred pH:
After cation column.
Before cation column.
N/A

Table of Contents

Other manuals for ABB AWT420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB AWT420 and is the answer not in the manual?

ABB AWT420 Specifications

General IconGeneral
BrandABB
ModelAWT420
CategoryTransmitter
LanguageEnglish

Related product manuals