AWT420 | UNIVERSAL 4-WIRE, DUAL-INPUT TRANSMITTER | OI/AWT420-EN REV. B
71
Turbidity TSS sensor verification
Preparing the verification tool and locking the
sensor in place
Refer to operating instruction OI/ATS430 -EN.
1 At the Calibrate level, press the key:
The Calibrate menu is displayed:
2 Use the / keys to select the sensor to be calibrated,
and press the key to confirm selection.
The TSS calibration menu is displayed:
3 Press the / keys to select Sensor Verification and
press the key to confirm selection.
The Verification Value is displayed:
4 Use the key to edit the verification value.
Use the / keys to set the value and press the key
to confirm changes.
Press the and key to proceed to the next step.
5 Ensure the sensor is inserted in the verification tool and
press the key to start the verification routine.
The verification process screen is displayed:
On completion the result is displayed.
or
Procedure Failed
Note. The verification process can be canceled at any
time during the process by pressing the Abort key ( ).
Calibrate
Back
Select
S1 : Turb TSS
S2 : Turb TSS
Hold Outputs
Calibrate
Back
Select
Sensor Verication
Turbidity Cal
TSS Cal
TSS Manual Cal
Sample Collection
Collection Complete
Sensor Verication
Verication Value
4000 NTU
Next
Edit
Verication
Start Verication
4000 NTU
Abort
Continue
Temperature Cal
NTU 3985 NTU
STD 4000 NTU
Settling - Please Wait
Abort
Procedure Pass

Table of Contents