EasyManuals Logo
Home>Agilent Technologies>Inverter>33250A

Agilent Technologies 33250A User Manual

Agilent Technologies 33250A
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
194
Chapter 4 Remote Interface Reference
Burst Mode Commands
4
BURSt:GATE:POLarity {NORMal|INVerted}
BURSt:GATE:POLarity?
Select whether the function generator uses true-high or true-low logic
levels on the rear-panel Trig In connector for an externally-gated burst.
The default is NORM (true-high logic). The :POL? query returns
“NORM” or “INV”.
OUTPut:TRIGger:SLOPe {POSitive|NEGative}
OUTPut:TRIGger:SLOPe?
Select a rising or falling edge for the “trigger out” signal. When enabled
using the OUTP:TRIG command (see below), a TTL-compatible square
waveform
with the specified edge is output from the rear-panel
Trig Out
connecto
r at the beginning of the burst. Select “POS” to output a pulse
with a rising edge or select “NEG” to output a pulse with a falling edge.
The default is POS. The :SLOP? query returns “POS” or “NEG”.
•When the Immediate (internal) trigger source is selected
(TRIG:SOUR IMM command), the function generator outputs a square
waveform with a 50% duty cycle from the Trig Out connector at the
beginning of the burst. The frequency of the waveform is equal to the
specified burst period (BURS:INT:PER command).
•When the External trigger source is selected (TRIG:SOUR EXT
command)
or when the Gated mode is selected (BURS:MODE GAT
command), the function generator automatically disables the
“trigger out” signal. The rear-panel Trig Out connector cannot be
used for both operations at the same time (an externally-triggered
waveform uses the same connector to trigger the burst).
•When the Bus (software) trigger source is selected (TRIG:SOUR BUS
command), the function generator outputs a pulse (>1 µs pulse width)
from the Trig Out connector at the beginning of each burst.
OUTPut:TRIGger {OFF|ON}
OUTPut:TRIGger?
Disable or enable the “trigger out” signal (used with burst and sweep only
).
When enabled, a TTL-compatible square waveform with the specified
edge (OUTP:TRIG:SLOP command) is output from the rear-panel
Trig Out connector at the beginning of the burst. The default is OFF.
The :TRIG? query returns “0” (OFF) or “1” (ON).

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals