Chapter 3 Features and Functions
Output Configuration
3
Function Limitations: If you change to a function whose maximum
frequency is less than that of the current function, the frequency is
adjusted to the maximum value for the new function. For example,
if you are currently outputting an 80 MHz sine wave and then change
to the ramp function, the function generator will automatically
adjust
the output frequency to 1 MHz (the upper limit for ramps).
Amplitude Limitations: If you change to a function whose maximum
amplitude is less than that of the current function, the amplitude is
automatically adjusted to the maximum value for the new function.
This may occur when the output units are Vrms or dBm due to the
differences in crest factor for the various output functions.
For example, if you output a 5 Vrms square wave (into 50 ohms) and
then change to the sine wave function, the function generator will
automatically adjust the output amplitude to 3.536 Vrms (the upper
limit for sine in Vrms).
Front-Panel Operation: To select a function, press any key in the top
row of function keys. Press to output the arbitrary waveform
currently selected. To view the other arbitrary waveform choices,
press the Select Wform softkey.
To select dc volts from the front panel, press and then select the
DC On softkey. Press the Offset softkey to enter the desired offset
voltage level.
Remote Interface Operation:
FUNCtion:SHAPe {SINusoid|SQUare|RAMP|PULSe|NOISe|DC|USER}
You can also use the
APPLy
command to select the function, frequency
,
amplitude, and offset with a single command.

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals