Chapter 5 Error Messages
Calibration Errors
4
5
Calibration Errors
The following errors indicate failures that may occur during a calibration
.
For more information on the calibration procedures, see chapter 4 in the
Agilent 33250A Service Guide.
701 Calibration error; security defeated by hardware jumper
The function generator’s calibration security feature has been disabled
due to the installation of a hardware jumper on the internal circuit board.
702 Calibration error; calibration memory is secured
A calibration cannot be performed when calibration memory is secured.
To unsecure the instrument, use the CAL:SEC:STAT ON command with
the correct security code.
703 Calibration error; secure code provided was invalid
The security code specified with the CAL:SEC:STAT ON command
is invalid.
705 Calibration error; calibration aborted
The function generator will not accept configuration commands (such as
APPL:SIN) while performing a calibration.
706 Calibration error; provided value is out of range
The calibration value specified with the
CAL:VAL
command is out of range.
707 Calibration error; signal input is out of range
The internal analog-to-digital converter (ADC) has determined that the
signal applied to the rear-panel Modulation In connector is out of range.
850 Calibration error; setup is invalid
You have specified an invalid calibration setup number with the
CAL:SET command. Refer to the Agilent 33250A Service Guide for more
information on the calibration procedures.
851 Calibration error; setup is out of order
Certain calibration setups must be performed in a specific sequence in
order to be valid. Refer to the Agilent 33250A Service Guide for more
information on the calibration procedures.