218
Chapter 4 Remote Interface Reference
Interface Configuration Commands
4
Interface Configuration Commands
See also “Remote Interface Configuration” on page 118 in chapter 3.
SYSTem:INTerface {GPIB|RS232}
Select the remote interface. Only one interface can be enabled at a time.
The GPIB interface is selected when the function generator is shipped
from the factory. The command does not have a query form.
SYSTem:LOCal
Place the function generator in the local mode for RS-232 operation.
All keys on the front panel are fully functional while in the local mode.
SYSTem:RWLock
Place the function generator in the remote mode for RS-232 operation
and disable all front-panel keys.
<Break>
Clear the operation in progress over the RS-232 interface and discard
any pending output data. This is equivalent to the IEEE-488 Device Clear
operation over the GPIB interface.