Chapter 4 Remote Interface Reference
The SCPI Status System
4
The Standard event register is cleared when:
You execute the *CLS command.
You query the event register using the *ESR? command.
The Standard Event enable register is cleared when:
You execute the *ESE 0 command.
You turn the power on and have previously configured the function
generator to clear the enable register using the *PSC 1 command.
Note that the enable register will not be cleared at power-on if you
have configured the function generator using the *PSC 0 command.