EasyManuals Logo
Home>Agilent Technologies>Inverter>33250A

Agilent Technologies 33250A User Manual

Agilent Technologies 33250A
336 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #251 background imageLoading...
Page #251 background image
251
Chapter 5 Error Messages
Command Errors
4
5
-158 String data not allowed
A character string was received but is not allowed for this command.
Check the list of parameters to verify that you have used a valid
parameter type. Example: BURS:NCYC ’TEN’
-161 Invalid block data
Applies only to the
DATA:DAC VOLATILE
command
.
For a definite-length block, the number of bytes of data sent does not
match the number of bytes that you specified in the block header. For an
indefinite-length block, an EOI (End-or-Identify) was received without
an accompanying <new line> character.
-168 Block data not allowed
Data was sent to the function generator in definite-length block format
but this command does not accept this format. Check to see if you have
sent the correct data type with the command. Example: BURS:NCYC #10
-170 to -178 Expression errors
The function generator does not accept mathematical expressions.

Table of Contents

Other manuals for Agilent Technologies 33250A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33250A and is the answer not in the manual?

Agilent Technologies 33250A Specifications

General IconGeneral
BrandAgilent Technologies
Model33250A
CategoryInverter
LanguageEnglish

Related product manuals