253
Chapter 5 Error Messages
Execution Errors
4
5
-221 Settings conflict;
burst count reduced
Since the burst period is currently at its maximum, the function generator
has reduced the burst count to allow the specified waveform frequency.
-221 Settings conflict;
trigger delay reduced to fit entire burst
The function generator has reduced the trigger delay in order to maintain
the current burst period and burst count
. The trigger delay is the time
between the receipt of the trigger and the start of the burst waveform.
-221 Settings conflict;
triggered burst not available for noise
You cannot use the noise function in the triggered burst mode. Noise is
allowed only in the gated burst mode.
-221 Settings conflict;
amplitude units changed to Vpp due to high-Z load
The output units (VOLT:UNIT command) cannot be set to dBm if the
output termination is currently set to “high impedance” (OUTP:LOAD
command). The function generator has converted the units to Vpp.
-221 Settings conflict;
trigger output disabled by trigger external
When the
external
trigger source is selected (
TRIG:SOUR EXT
command
),
the function generator automatically disables the “trigger out” signal.
The rear-panel Trig connector cannot be used for both operations at the
same time (an externally-triggered waveform uses the same connector to
trigger the sweep or burst).
-221 Settings conflict;
trigger output connector used by FSK
If you have enabled FSK and have selected the external trigger source
(FSK:SOUR EXT command), the “trigger out” signal cannot be enabled
(OUTP:TRIG ON command). The rear-panel Trig connector cannot be
used for both operations at the same time.