Performance Tests and Calibration 6
Agilent U1251B/U1252B User’s and Service Guide 137
Calibration Process
To complete a full instrument calibration use following
general procedures:
1 Read “Test Considerations” on page 124.
2 Perform the verification tests to characterize the
instrument (incoming data).
3 Unsecure the instrument for calibration (see “Calibration
Security” on page 125).
4 Perform the adjustment procedures (see “Adjustments
Consideration” on page 139).
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the
instrument's maintenance records.
Make sure to quit the Adjustment Mode before turning off the instrument.