162 Agilent U1251B/U1252B User’s and Service Guide
7 Specifications
Capacitance Specifications
Temperature Specifications
Tab l e 7-8 Capacitance specifications
Range Resolution
Accuracy
± (% of reading + Offset
Error)
Display update rate (approx)
10.000 nF 0.001 nF 1% + 8
4 times/sec.
100.00 nF 0.01 nF
1% + 5
1000.0 nF 0.1 nF
10.000 µF 0.001 µF
100.00 µF 0.01 µF
1000.0 µF 0.1 µF 1 time/sec.
10.000 mF 0.001 mF 0.1 times/sec.
100.00 mF 0.01 mF 3% + 10 0.01 times/sec.
Notes for capacitance specifications:
1 Use the null operation to zero out residual offset before measuring the signal (by opening the test leads).
Tab l e 7-9 Temperature specifications
Thermocouple
Typ e
Range Resolution
Accuracy
± (% of reading + No. of
Least Significant Digit)
K
–200 – 1372 °C/
–328 – 2502 °F
0.1 °C/
0.1 °F
0.3% + 3 °C/
0.3% + 6 °F
J
[2]
–210 – 1200 °C/
–346 – 2192 °F
0.1 °C/
0.1 °F
0.3% + 3 °C/
0.3% + 6 °F

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1252B and is the answer not in the manual?

Agilent Technologies U1252B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1252B
CategoryMultimeter
LanguageEnglish

Related product manuals