2 Operations and Features
44 U3402A User’s and Service Guide
Procedure
1 Press to enable Rel operation.
2 Press
to toggle in (and out) the Rel #.
3 Use , , , and to modify the Rel #.
4 Press to set the Rel # and read the display.
5 Press to disable the Rel operation.
MinMax
The MinMax (Minimum/Maximum) operation stores the minimum and
maximum values of reading during a series of measurements.
When enabled, the MinMax operation turns on the MINMAX annunciator and
begins accumulating various statistics of the readings being displayed.
Figure 2-15 Typical Max operation display
• In resistance measurement mode, the multimeter will read a non-zero value even when
the two test leads are in direct contact, because of the resistance of these leads. Use
the Rel operation to zero-adjust the display.
• In DC voltage measurement mode, the thermal effect will influence the accuracy. Short
the test leads and press once the displayed value is stable to zero-adjust the
display.