Performance Test 4
U3402A User’s and Service Guide 71
Diode Verification Test
Configuration: Diode
1 Connect the calibrator to the front panel Hi and Lo input terminals.
2 Select each function and range in the order shown in Table 4- 6. Provide the
input shown in Table 4- 6.
3 Make a measurement and observe the result. Compare measurement results
to the appropriate test limits shown in Table 4- 6. (Be certain to allow for
appropriate source settling when using the Fluke 5520A.)
Frequency Verification Test
Configuration: Frequency
1 Select the frequency function.
2 Select each range in the order shown in Table 4- 7. Provide the input voltage
and frequency indicated. Compare measurement results to the appropriate
test limits shown in Table 4- 7. (Be certain to allow for appropriate source
settling.)
Ta b l e 4- 6 Diode verification test
Function Reading rate Voltage Range Error from nominal
one year
Diode
Slow 0.50000 V 1.2 V ±0.11 mV
1.00000 V 1.2 V ±0.17 mV
Medium 0.5000 V 2.5 V ±0.5 mV
2.0000 V 2.5 V ±0.7 mV
Ta b l e 4- 7 Frequency verification test
Function Reading
rate
Voltage Input
frequency
Range Error from nominal
one year
Frequency Slow 1 V 1000.00 Hz 1200 Hz ±0.08 Hz
U3402-90001.book Page 71 Friday, July 24, 2009 4:04 PM