EasyManuals Logo
Home>Commtest Instruments>Measuring Instruments>vb2000

Commtest Instruments vb2000 User Manual

Default Icon
203 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #190 background imageLoading...
Page #190 background image
Appendix 2: Specifications
181
Appendix 2: Specifications
Specifications Model vb3000 Remarks
Sensor Input
Number of channels
Accelerometers
Velocity sensors
Displacement sensors
Current sensors
Connector
Input impedance
Voltage swing
Sensor excitation current
Sensor excitation voltage
Sensor detection
2
2-wire, low impedance piezoelectric, 100 mV/g nominal
100 mV/in/s (4 mV/mm/s) nominal
100 mV/mil (4 mV/µm) nominal
100 mV/amp nominal
BNC
>100 kΩ
16 V peak-peak
0 mA or 2.2 mA (configurable)
24 V maximum
Warns if short circuit or not connected
Commonly termed 'ICP® type'. Sensitivity (8.5 to 2300) mV/g
Sensitivity (8.5 to 2300) mV/in/s = (0.34 to 90.55) mV/mm/s
Sensitivity (8.5 to 2300) mV/mil = (0.34 to 90.55) mV/µm
Sensitivity (0.1 to 2300) mV/amp
Safety feature: Break-free inline connector
AC coupled input, allows for ± 8 V sensor output swing (± 80 g)
2.2 mA required for ICP® type accelerometer
At sensor terminals with sensor attached
Channel 1 only
Tachometer Input
Sensor
Laser sensor range
Sensor supply
Input type
Pulse rating
Speed range
Display
Laser sensor with reflective tape included in kit
10 cm to 2 m nominal
7.2 V nominal 6.0 V to 9.5 V instrument battery
Optically isolated, accepts TTL pulse
2.5 V (4 mA) min, 10 V (27 mA) max, off-state < 0.8 V
30 RPM to 65 000 RPM (0.5 Hz to 1.08 kHz)
RPM, Hz, 1X amplitude and phase angle
Sensor triggers when the tape reflects its beam
Dependent on size of reflective tape
Available to power sensor. Protected by 0.1 A PTC
Triggers on negative edge
For selected amplitude type, phase angle in degrees
Parameter Indication
Displays
Maximum levels
Dynamic signal range
Harmonic distortion
Units
Graph types
Magnitude display
Warnings
Cursors
Accuracy
Frequency response
Acceleration, velocity, displacement, demodulation and
current
± 80 g (800 m/s
2
), ± 4 in/sec (100 mm/s), ± 800 amps
± 400 mil (10 mm)
95 dB (typical at 400 line resolution)
Less than -70 dB typical
g or m/s
2
, in/s or mm/s, mil or mm or µm, amps
AdB, VdB
Spectrum (freq domain), waveform (time domain)
Overall rms value, cursor-position value
% change in overall since baseline
Standard cursor
Dual cursors
Harmonic cursor
± 1% (0.1 dB)
± 0.1 dB from 10 Hz to 15 kHz;
± 3 dB from 1 Hz to 40 kHz
User selectable
-peak. Typical, dependent on individual calibration
Acceleration and velocity. Greater with higher resolution and
averaging
Dependent on input level and type. Other distortions and noise
are lower
-peak, peak-peak or rms
AdB ref. 1 µg rms, VdB ref. configurable 1.0e-5 mm/s rms or 1.0e-
6 mm/s rms
Solid histogram for spectrum, line graph for waveform
Digital readout on chart
Tolerances: Tight 50% to 150%, relaxed 25% to 200%
Vary x position to display x and y values
Lock standard cursor as reference and display difference
Up to 32 whole-number multiples of standard-cursor frequency
Measured at 100 Hz, 23 ± 5 °C, 400 lines, 400 Hz range
Acceleration and velocity. From value measured at 100 Hz
Spectrum Display
Fmax possible ranges
Fmin possible range
Resolution
Frequency scale
Amplitude scale
Window shapes
Overlap
Number of averages
Averaging types
Demod bandwidths
0 to (100, 125, 150, 200, 300, 400, 500, 600, 800) Hz
0 to (1, 1.2, 1.6, 2, 2.5, 3, 4, 5, 6, 8, 10, 15, 20, 30, 40) kHz
0 to Fmax
400, 800, 1600, 3200, 6400 lines (configurable)
Hz, CPM, orders
Acceleration, velocity, displacement or current
Hanning, rectangular
(0, 12.5, 25, 37.5, 50, 62.5, 75, 87.5) %
1, 2, 4, 8, 16, 32, 64, 128
Linear, exponential, peak hold, synchronous
20 bandwidth options
Or equivalent CPM values
Or orders-based from 1X to 30 000X
vb instrument zeroes all spectral lines below Fmin
3200 lines maximum for dual channel measurements
Linear scale. Can zoom in to display individual spectral lines
Linear or log scales
Dependent on Fmax and number of lines
Increases sampling time proportionally
From 125 Hz to 1250 Hz up to 16 kHz to 20 kHz
Waveform Display
Number of samples
Long time waveform
Time scale
Time synchronous averages
1024, 2078, 4096, 8192, 16 384
Up to 15 kHz Fmax. Over 17 hours duration or until memory
full
10 ms to 32 seconds/ 1 rev to 30 000 revs
1, 2, 4, 8, 16, 32, 64, 128
Only available when tachometer triggered
Keypad Entry
Prompt and unit strings
Input value range
16 characters each
± 59 999
Time Intervals
Lines
Measuring time in seconds
(example ranges)
Range
0 Hz to100 Hz
0 Hz to 800 Hz
0 Hz to 4 kHz
0 Hz to 40 kHz
400
4
0.5
0.1
0.01
800
8
1
0.2
0.02
1600
16
2
0.4
0.04
3200
32
4
0.8
0.08
6400
64
8
1.6
0.16
Dependent on number of lines and number of averages (values
shown in table for no overlap, no averaging, maximum display
update rate of 4 per second)
Typical recording time 5 s for 1600 lines, 1600 Hz, 8 averages, 50% overlap Not including initial startup and settling time
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Commtest Instruments vb2000 and is the answer not in the manual?

Commtest Instruments vb2000 Specifications

General IconGeneral
BrandCommtest Instruments
Modelvb2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals