The maximum and minimum must be reset individually.
7.8 Reset meter to factory defaults
Reset to factory defaults will restore most meter parameters to their factory default
setting.
When the parameters have been changed in a way that is difficult to determine what’s
happening, it might be better to start the setup process from the factory defaults.
Factory default reset does not change the saved backup restore settings, override
passwords, or edit parameters locked by a permanent non-resettable grand total. See
Section 6.10.4.
Instructions to load factory defaults can be found in the Section 6.12.5 menu.
7.9 Factory defaults and user settings
The following table shows the factory setting for most of the programmable parameters
on the meter. Next to the factory setting, the user may record the new setting for the
particular application.
Default and user settingsTable 7-3:
Model: S/N: Date:
Parameter Display Default setting User setting
Input type InPUt Active
K-Factor units Funit Pulse/gallon
F-Factor FActr 1.0000
Rate time base tbASE Second
Rate unit rAtEU Gallons/second
Total unit tot U Gallons
Total multiplier muLt x1
Grand total unit GtotU Gallons
Grand total multiplier muLt x1
Rate decimal point 1111.1 1 place
Total decimal point 11111.1 1 place
Grand total decimal
point
11111.1 1 place
Total conversion factor totCF N/A (only valid with
customs units)
Grand total conversion
factor
GrtCF N/A (Only valid with
custom units)
Startup procedure
User manual 91