6.12.6 Battery power symbol alert (BAT SyM)
The Battery Power Symbol Alert menu is used to enable or disable the battery alert symbol
on the display. This is a useful way to be aware of a power failure to a model with battery
backup.
When enabled, the battery symbol will appear whenever the meter is powered as a battery
backup. This is detected when the meter being powered from DC or loop-power
experiences power loss, subsequently switching over to battery power.
The indicator will not appear if the meter is powered on via battery power, only when there
is applied power to the DC or loop-power lines, followed by power loss. This prevents the
batter indicator from appearing at all times for a primarily battery powered application.
The battery symbol will flash in a low battery condition regardless of the setting of this
parameter.
6.12.7 Information (INFO)
The Information menu is part of the System features menu. It shows software
identification number, version number, and extended model number. To view this
information:
1. Go to the Information menu (INFO) and press Enter button.
2. Continue pressing Enter to scroll through the displays.
3. Follow the information display to exit the meter's Advanced features menu and
return to Run mode.
6.13
Serial communications (COMM)
The Communications menu is used to setup serial communications parameters necessary
for communication via Modbus.
Product advanced features