EasyManuals Logo

Fresenius Medical Care 2008K Calibration Manual

Fresenius Medical Care 2008K
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Page 32
Fresenius 2008
®
K Calibration Procedures
P/N 507296 Rev D
2.3.2 VENOUS PRESSURE CALIBRATION
From the Calibrate Sensors screen, select the Venous Pressure screen button.
The screen will change to the following:
1. Open the venous transducer port P
VEN
.
on the level detector module to air
(atmospheric pressure).
2. Adjust the ZERO SET potentiometer (see Figure 6) so that the Venous
Pressure data box indicates zero mmHg.
Press the [CONFIRM] key and the screen will change.
3. Attach a syringe and a calibrated pressure gauge to the P
VEN
.
port using a
T- fitting.

Other manuals for Fresenius Medical Care 2008K

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K and is the answer not in the manual?

Fresenius Medical Care 2008K Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008K
CategoryMedical Equipment
LanguageEnglish

Related product manuals