Page 57
Fresenius 2008
®
K Calibration Procedures
P/N 507296 Rev D
2.3.8 CONDUCTIVITY CELLS CALIBRATION
If the machine is equipped with Functional Software v2.16 or less then perform
Section 2.3.8.1. If the machine is equipped with Functional Software v2.25 or
greater then perform Section 2.3.8.2.
2.3.8.1 CONDUCTIVITY CELLS CALIBRATION (v2.16 or less)
Machines equipped with Online Clearance (OLC) have two conductivity cells.
The first conductivity cell is PRE dialyzer and the second conductivity cell is
POST dialyzer.
Note: If the Clearance option is set to NO in Hardware Options, the post
conductivity cell will not be calibrated.
From the Calibrate Sensors screen, select the Cond Cells screen button. The
screen will change to the following: