Page 41
Fresenius 2008
®
K Calibration Procedures
P/N 507296 Rev D
2.3.6 TEMPERATURE CONTROL CALIBRATION
The Temperature Control Calibration is dependent upon Function Software
version and enabled options. Perform the appropriate calibration according to the
list below:
Function Software V2.16 or less:
Perform section 2.3.6.1 if all the following apply:
- Slow Flow option is disabled.
Perform section 2.3.6.2 if all the following apply:
- Slow Flow option is enabled
Function Software V2.25 or greater:
Perform section 2.3.6.1 if all the following apply:
- OLC option not installed OR OLC option is installed but Temp Comp
option is disabled.
- Slow Flow option is disabled.
Perform section 2.3.6.2 if all the following apply:
- OLC option not installed OR OLC option is installed but Temp Comp
option is disabled.
- Slow Flow option is enabled.
Perform section 2.3.6.3 if all the following apply:
- OLC option is installed and Temp Comp option is enabled.
- Slow Flow option is disabled.
Perform section 2.3.6.4 if all the following apply:
- OLC option is installed and Temp Comp option is enabled.
- Slow Flow option is enabled.

Other manuals for Fresenius Medical Care 2008K

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K and is the answer not in the manual?

Fresenius Medical Care 2008K Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008K
CategoryMedical Equipment
LanguageEnglish

Related product manuals