Page 49
Fresenius 2008
®
K Calibration Procedures
P/N 507296 Rev D
2.3.6.3 TEMPERATURE CONTROL CALIBRATION (METHOD 3)
Perform this section only if the machine is equipped with the following:
- Functional Software v2.25 or greater, the machine has the OLC option
installed AND the Temp Comp option is enabled and the slow flow option is
disabled.
Note: The Temp Comp option is only present in Functional Software v2.25 or
greater and uses post temperature sensor NTC44. Do not attempt to calibrate a
non-OLC machine with the Temp Comp option enabled since NTC44 is not
present and an Operator Error will occur.
Replace any panels that have been removed so that the machine is closed up as it
would be in normal operation. It is important for this calibration that ventilation
around the hydraulics unit be the same as it is when the machine is used for
dialysis.
From the Calibrate Sensors screen, select the Temp Control screen button.
The screen will change to the following:

Other manuals for Fresenius Medical Care 2008K

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K and is the answer not in the manual?

Fresenius Medical Care 2008K Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008K
CategoryMedical Equipment
LanguageEnglish

Related product manuals