Page 3
Fresenius 2008
®
K Calibration Procedures
P/N 507296 Rev D
Digital Multimeter to measure dc voltage, ac voltage and resistance with an
accuracy of at least 1% + 1 digit for dc voltage measurements 1.5% + 5
digits for ac measurements and 1% + 1 digit for resistance measurements.
Syringe, 60 cc capacity. Tolerance is not important; the syringe is not used
for volume measurements.
For machines equipped with the blood pressure module, the following equipment
is also required to test the module as described in this manual:
Test Device (Fresenius part number 370090). The Test Device contains two
air chambers with calibrated volumes.
Mercury manometer or equivalent pressure meter accurate to within 1
mmHg at pressures up to 335 mmHg.
1.2 OPERATING MODES
The following calibration procedures contain instructions to place the 2008K into
Dialysis Mode and Service Mode.
To place the machine in Service Mode, turn the machine power On and wait for
the message Press CONFIRM for Service Mode to appear. Once it appears,
press the [CONFIRM] key and the message will change to Machine in Service
Mode. After the System Initializing process is complete, the machine will be in
Service Mode.
If the [CONFIRM] key is not pressed when the Press CONFIRM for Service
Mode message is on the screen, the screen will change and the message Machine
in Dialysis Mode will appear. After the System Initializing process is complete,
the machine will be in Dialysis Mode.