Page #41 background imageLoading...
Page #41 background image
3 Operation
K0443 Revision A 3 - 16 [EN] English
3.8 Supervisor set-up
enter four digit code
(0268)
Enable/disable
-
-
-
Protective Vent
In Limits
Alarms
Communications
Time out
Restore last settings
Supervisor
More
next page:-
PIN
Pressing the escape key stores
settings and returns to global
set-up.
High source pressure
Low source pressure
Enable/disable high pressure
Enable/disable low pressure
IEEE488
RS232
USB
Ethernet
% full-scale
Time period - time to stay in limits before
trigger indication
Idle time out - Enable/disable
Gas head correction
Enable/
disable
Head - in or m
Local gravity - m/s
2
Barometric - mbar
Reference port medium*
Test port medium*
Temperature (gas)
Humidity (gas)
*air or nitrogen
Power-up
Normal
Regulator - Power-on set-point
Lock
Lock/unlock individual tasks
Lock/unlock all tasks
Lock tasks
Change PIN
Enter new PIN
Re-enter new PIN
Enter new
supervisor PIN
User defined units
Enter the name of the defined units
Enter the equivalent value in Pascals
User defined units -
user unit 1
Instrument alias name
Language
More
next page:-
Restore previous defaults Y/N?

Table of Contents

Other manuals for GE PACE6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PACE6000 and is the answer not in the manual?

GE PACE6000 Specifications

General IconGeneral
BrandGE
ModelPACE6000
CategoryTest Equipment
LanguageEnglish

Related product manuals