6 Reference and Specification
K0443 Revision A 6 - 24 [EN] English
6.8 Options
Option enable process
To enable soft options on a PACE instrument, use the following:
1. Touch the top Measure area of the
screen.
2. Select Global Setup.
3. Select Calibration.
4. Enter a Calibration PIN 1234.
5. Enter new option key xxxxxxxxxx (10
digits).
6. After entry of this key PACE confirms
the options have been enabled.
Options in the Task Menu
The following software enabled options
can be selected from the task menu:
Leak Test option
This task applies one or two test pressures to either an external system to find any leaks in a
system connected to the instrument or an internal leak check. This task sets the test
pressure, control dwell time at the test pressure and the leak test time (measure dwell time).
Serial
number

Table of Contents