Installation

Model
1630A/D/G
LINE
VOLTAGE
SELECTION
2-2
BEFORE
CONNECTING
THIS
INSTRUMENT
to
line
(Mains)
voltage,
be
sure
the
line
voltage
switch
is
set
correctly
and
that
the
proper
fuse
is
installed
.
If
the
line
fuse
burns
out,
do
not
replace
it
until
the
cause
for
failure
has
been
determined
and
repaired
by
a
qualified
service
person
.
Replacing
this
fuse
in
a
damaged
instrument
can
cause
additional
damage
.
AC
POWER
CABLE
OPERATOR'S
MAINTENANCE
WARNING
BEFORE
CONNECTING
THIS
INSTRUMENT,
the
protective
earth
terminal
of
the
instrument
must
be
connected
to
the
protective
conductor
of
the (Mains)
power
cord
.
The
Mains
plug
must
be
inserted
in
a
socket
outlet
provided
with
a
protective
earth
contact
.
The
protective
action
must
not be
negated
by
the use
of
an
extension
cord
(power
cable)
without a
protective
conductor
(grounding)
.
Grounding
one
conductor
of
a
two conductor
outlet
does
not
provide
an instrument
ground
.
The
only
instrument
maintenance
an
operator
needs
to
perform
is
to
occasionally
clean
the
display
screen
and
the
front
panel
keyboard
.
When
cleaning
the
1630A/D/G,
caution
must
be
used
with
which
cleaning
agents
are
used
.
Use
a
mild
soap
and
water
solution
.
If
a harsh
soap
or
solvent
is
used,
the
water-base
paint
finish
will
be
damaged
.
SITE
SELECTION
REAR
PANEL
ADDRESS
SWITCH
CAUTION
Use
care
when
cleaning
the
front
panel
keyboard
.
Water
can
damage
the
keyboard
circuitry
if
it
seeps
under
the
keys
.
Do
not
use
chemical
cleaning
agents
or
abrasive cleaners
that
may
damage
the
plastics
in
this
instrument
.
Recommended
cleaning
agents
are
isopropyl
alcohol,
kelite
(1
part
kelite,
20
parts
water),
or
a
solution
of
1%
mild
detergent
and
99%
water
.
Place
the
logic
analyzer
on
a
clean
workspace
which
has
adequate
ventilation
.
The
dip
(dual-inline-package)
switch
on
the
rear
panel
of
the
logic
analyzer
is
used
to
configure
the
analyzer
for
operation
on
the
HP-IB
and/or
HP-IL
interface
buses,
and
to
execute
the
self-test
routine
.
The
first
five
switches
set
the
address
of
the
analyzer
for
HP-113
operation
.
Switch
six
executes
the
analyzer
self-test
when
set
to
"1"
.
The
positions
of
switches
7
and
S
determi
ne
the
analyzer
configuration
on
the HP-I
B/HP-I
L
interface
buses
.

Table of Contents

Other manuals for HP 1630D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 1630D and is the answer not in the manual?

HP 1630D Specifications

General IconGeneral
BrandHP
Model1630D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals