Model
1630A/D/G

Timing
Measurements
Controlling
The
Timing
Display
1
.
Move
the cursor
to
the
Magnification
field
in
the
Timing
Waveform
Diagram
.
2
.
Press
the
NEXT[]
key
to
obtain
4X
magnification
.
This
shows
a
window
in
the
timing
memory
.
One-
fourth
of
the
memory
is
on
screen
.
3
.
Press
the
ROLL
down
key
to
see
the
window
move
through
the
timing
memory
(display
moves
across
the
screen)
.
4
.
Press
the
NEXT[]
key
several
times
to
see
the magnifications
available
.
Select
a
magnification
factor
that
leaves several
waveform
transitions
on
screen
.
5
.
Press
and
hold
the
CURSOR
right
key
until
the
"x"
marker
moves
out
onto
the
display
area
.
Place
the
"x"
marker
on
one
of
the
waveform
transitions
on
the
display
.
The
first
line
above
the
waveforms
on
the
right-hand
side
of
the
screen
shows
the
measurement
of
time
between
the
start
of
the
memory
and
the
position
where
you
placed
the
"x" (o
to x)
.
6
.
Press
the
NEXT[]
key
several
times
to
see
that
magnification
of
the
display
occurs
around
the
area
in
memory
you
selected
by
the
positon
of 'Y'
.
See
figure
6-3
.
Sample Period
Magnification
Magnify
About
Cursor
Mopes
[
1
1
0
Timing
Waveform
Diagram
----------
Data
Acquired
Oct 27 1982
16t46
-1-
100
.0
nS/dig
10
.00
nS/clk
6
.300
pS o to
x
Figure
6-3
.
Magnification
Around
"x"
7
.
Press
the
CURSOR
down
key
to
move
the
cursor
to
the
Magnify
About
field
.
8
.
Press
the
NEXT[]
key
to
change
"x" to
"o"
.
9
.
Move
the
cursor
up
into
the
Magnification
field
.
10
.
Press
the
NEXT[]
key
several
times
.
Notice
that
display
magnification
now
occurs
around
the
position
of
"o"
(at
the
start
of
the timing
memory)
.
11
.
Select
a
magnification
that
leaves
the
"x"
marker
on
screen
.
12
.
Press
the
CURSOR
down
key
to
move
the
cursor
to
the
Cursor
Moves
field
.
13
.
Press
the
NEXT[]
key
to
change
"x"
to
"o"
.
6-3

Table of Contents

Other manuals for HP 1630D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 1630D and is the answer not in the manual?

HP 1630D Specifications

General IconGeneral
BrandHP
Model1630D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals