HP 54753A TDR/TDT Setup Menu
TDR 1 dest
4-5
If no other valid TDT destination channels are available, then "none" is the only
choice. If a TDT destination other than none is selected, then the
Preset
TDR/TDT
control will turn on and preset the TDT destination channel.
Key Path TDR/TDT Setup
TDT 1 dest
TDR 1 dest
The
TDR 1 dest
softkey only appears when the stimulus is set to external.
Pressing the
TDR 1 dest
softkey produces a pull-down menu used to select the
channel used as the destination channel for TDR measurements. Any electrical
channel is a valid TDR destination channel. The TDR destination may not be
set to the currently defined TDT destination.
Key Path TDR/TDT Setup
TDR 1 dest
Normalize response . . .
The
Normalize response
function allows you to change the risetime of the
normalized step, to select TDR and TDT normalization, to turn on or off the
display of the normalized TDR or TDT trace, to change the scaling of the
normalized trace, and to establish the normalization filter values and reference
plane.
Risetime
The Risetime function allows you to change the normalized step’s risetime from
a minimum of
10 ps
or
whichever is greater, to a maximum of
min 8 points
time per division (s/div) 10 divisions×
record length
---------------------------------------------------------------------------------------------
×=
max 5 time per division (s/div)×=