EasyManuals Logo
Home>HP>Control Unit>54753A

HP 54753A User Manual

HP 54753A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
HP 54754A TDR/TDT Setup Menu
Preset TDR/TDT
5-12
Preset TDR/TDT
The Preset TDR/TDT function prepares the oscilloscope for making TDR/TDT
measurements by automatically setting several menu fields. The TDR preset
feature appears in the TDR/TDT Setup menu once a stimulus has been selected.
The menus that are affect by this feature are shown in Table 5-3.
Table 5-3
Preset TDR/TDT Configuration
Key Path TDR/TDT Setup
Preset TDR/TDT
Menu Menu Item Set To
Acquisition Averaging on
Best Flatness
Time Base Scale 500.0 ps/div
Position Set to a value which places the
incident edge on screen.
Channel 1 Setup Display on
Scale 100 mV/div
Offset 200.0 mV
Bandwidth 12.4 GHz
Attenuation units ratio
Attenuation 1.000 : 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 54753A and is the answer not in the manual?

HP 54753A Specifications

General IconGeneral
BrandHP
Model54753A
CategoryControl Unit
LanguageEnglish

Related product manuals