HP 54754A TDR/TDT Setup Menu
Preset TDR/TDT
5-22
rate varies automatically as the Time base
Scale
is changed to keep multiple
steps off screen. As the TDR rate decreases, TDR measurements can be made
on longer transmission lines.
Key Path TDR/TDT Setup
TDR rate
automatic
Key Path TDR/TDT Setup
TDR rate
manual
Preset TDR/TDT
The Preset TDR/TDT function prepares the oscilloscope for making TDR/TDT
measurements by automatically setting several menu fields. The TDR preset
feature appears in the TDR/TDT Setup menu once a stimulus has been selected.
The menus that are affect by this feature are shown in Table 5-7.
Table 5-7
Preset TDR/TDT Configuration
Key Path TDR/TDT Setup
Preset TDR/TDT
Menu Menu Item Set To
Acquisition Averaging on
Best Flatness
Time Base Scale 500.0 ps/div
Position Set to a value which places the
incident edge on screen.
Channel 1 Setup and Display on
Channel 2 Setup Scale 100 mV/div
Offset 200.0 mV
Bandwidth 12.4 GHz
Attenuation units ratio
Attenuation 1.000 : 1