HP 54754A TDR/TDT Setup Menu
TDR rate automatic . . . (250 kHz)
5-16
The function steps you through the normalization and reference plane
procedure for the selected measurement type. The procedure steps are
displayed at the top of the screen. The items required for calibration are shown
in Table 5-4.
Table 5-4
Establish normalization & ref plane Hardware Requirements
Key Path TDR/TDT Setup
Normalize response . . . Establish normalization & ref plane . . .
TDR rate automatic . . . (250 kHz)
The
TDR rate automatic . . . (250 kHz)
function allows you to manually or
automatically select the repetition rate of the TDR step. The range of values
for manual mode selection are from 50 Hz to 250 kHz repetition rate using a
1-2-5 sequence. When this function is set to automatic, the TDR step repetition
rate varies automatically as the Time base
Scale
is changed to keep multiple
steps off screen. As the TDR rate decreases, TDR measurements can be made
on longer transmission lines.
Key Path TDR/TDT Setup
TDR rate
automatic
Key Path TDR/TDT Setup
TDR rate
manual
Measurement Requirements
TDR 50 ohm 3.5 mm SMA terminator
3.5 mm SMA short
TDT 1 or 2 each 3.5 mm SMA cables
3.5 mm barrel connector